The SYCON 2602 is an analysis device for limit value monitoring. It enables a demand-controlled regeneration triggering of your softening system or the monitoring of a reverse osmosis. The limit value of the SYCON 2602 analyzer is determined by the indicator used. In addition to the current measured values, the display also shows other information such as the history of the last 100 measurements. In addition, all measurement results are documented on an SD card.
The limit is determined by the reagent used in the H25 or C25 series.
Specifications
Parameters | value / range |
Power supply | 24V AC/DC ±10% |
Power consumption | 25 VA (peak) |
Protection class | I |
degree of protection | IP 43 (housing version IP 54) |
Ambient temperature | 10 ... 45 °C |
Measuring water temperature | 5 ... 40 °C |
Sample port | Hose 6mm |
Dimensions (W x L x D) | 274x275x130mm |
Weight | 1.9kg |
Operating pressure | 0.5 .. 5 bar |
Reagent | H25 for residual / total hardness C25 for carbonate hardness / acid capacity |
Durability reagent | 24 months |
Article number | 30-010146 |
Article description | SYCON 2602 with housing (24 VAC/DC) |
We manufacture our reagentourselves. Only high-quality raw materials are used. Each production batch is checked in our laboratory for the specified limit value.
Reagent | Article number | Threshold | ||||
Single Bottle | Value Pack | °dH | °f | ppm CaCo3 | °e | |
H25-0.02 | 32-084115 | 32-484115 | 0.02 | 0.04 | 0.4 | 0.03 |
H25-0.05 | 32-084125 | 32-484125 | 0.05 | 0,1 | 1 | 0.06 |
H25-0.1 | 32-084135 | 32-484135 | 0,1 | 0.2 | 2 | 0.13 |
H25-0.2 | 32-084145 | 32-484145 | 0.2 | 0.4 | 4 | 0.25 |
H25-0.3 | 32-084155 | 32-484155 | 0.3 | 0.6 | 3 | 0.38 |
H25-0.5 | 32-084165 | 32-484165 | 0.5 | 1.0 | 10 | 0.63 |
H25-1 | 32-084175 | 32-484175 | 1 | 2.0 | 20 | 1.25 |
H25-2 | 32-084185 | 32-484185 | 2 | 4.0 | 40 | 2.5 |
H25-3 | 32-084195 | 32-484195 | 3 | 6.0 | 60 | 3,8 |
H25-5 | 32-084205 | 32-484205 | 5 | 10 | 100 | 6,3 |
H25-10 | 32-084215 | 32-484215 | 10 | 20 | 200 | 12.6 |
We manufacture our reagent ourselves. Only high-quality raw materials are used.Each production batch is checked in our laboratory for the specified limit value.
Reagent | Article number | Threshold | ||||
Single Bottle | Value Pack | °dH | °f | ppm CaCo3 | mmol/L (KS4.3) | |
C25-1 | 32-086125 | 32-486125 | 1 | 2 | 20 | 0.357 |
C25-1.5 | 32-084125 | 32-484125 | 1.5 | 3 | 30 | 0.535 |
C25-2 | 32-084135 | 32-484135 | 2 | 4 | 40 | 0.714 |
C25-3 | 32-084145 | 32-484145 | 3 | 6 | 60 | 1,071 |
We also offer accessories for our devices for easy installation and maintenance
Description | Article number | Description |
SYCON connection set | 33-000701 | For connecting a SYCON analyzer to a sample line with 1/2" or 1/4"
|
Pressure reducer with filter incl. manometer for SYCON |
33-090738 | Recommended for inlet pressures above 2 bar
|
Cleaning set SYCON Clean | 30-010900 | Cleaning set for measuring combs
|
Cleaning liquid FIT3000 | 32-089100 | Cleaner for measuring chambers (1000 ml) |
Maintenance set SYCON 2500 - SYCON 2602 |
32-089100 | Spare parts set for SYCON 2500 - 2602
|