Sycon 3100-EDTA

Print

The SYCON 3100-EDTA is an online analyzer for the fully automatic determination of EDTA. In addition to measuring the EDTA concentration, this model forms the basis for customized solutions.

Available from 3rd quarter 2019

 

The special features of the Sycon 3100-EDTA analyzer include:

  • easy commissioning
  • intuitive operation
  • low maintenance
  • self-calibrating
  • diagnostic program
  • 4...20 mA interface
  • 2 contact inputs for external control or current switch
  • maintenance-free full-colour sensor
  • tool-free maintenance of the measuring chamber


The following functions are reliably displayed by the Sycon 3100-EDTA analyzer:

  • free selectable limit values in the measuring range of the reagent used
  • alarm in case of lack of reagent / BOB functionality
  • graphic display (illuminated)
  • various languages
  • analysis startup options:
    • automatically by time interval
    • suppression of the analysis via current switch
    • analysis via external control


Article number 30-010168
Article description Online analyzer SYCON 3100-EDTA on wall mounting plate
Scope of delivery Analyzer, manual
Parameter Value / Range
Supply voltage 85 ... 264 VAC at 47 ... 440 Hz
Power consumption 25 VA (operating)
Protection class IP 43 - with housing IP 54
Ambient temperature 5 ... 45 °C
Temperature of measuring water 5 ... 45 °C
Humidity 20 ... 90 % RF not condensing

 

Parameter Value / Range
Installation wall mounting in closed rooms
Dimensions on wall holder:
274 x 275 x 130 mm
with housing (has to be ordered separately):
330 x 300 x 185 mm
Weight on wall holder: 1,9 kg
with housing: 3,7 kg
Relais outputs 4 relays
250 VAC / VDC 4 A
potential free outputs NC/NO
  • alarm "limit value"
  • dysfunction of analyzer
  • analysis active / cooling water
  • lack of reagent
Signal input potential free input contact
Supply intake/outlet

tubes with 6 mm outer diameter

 

 

 

Parameter Value / Range
Measuring principle titrimetrical method
Measuring range
EDTA unit
5 ... 50 ppm
Reagent consumption  
Shelf-life of reagent 24 months
Water consumption The water consumption varies depending on the inlet pressure and the flushing duration.
Inflow pressure approx. 0,5 ... 5 bar
If the inlet pressure exceeds 2 bar a pressure reducer has to be used.
Water sample clear, colourless, without solids, without gas bubbles
Parameter Value Unit
ph 4 ... 10,5   
iron < 3 ppm
copper < 0,2 ppm
aluminium < 0,1 ppm
manganese < 0,2 ppm
acid capacity KS 4.3 < 5 mmol/l

 

 