The following work steps refer to the measuring chamber of the analysers of the Sycon series:
- Observe the regulations for handling hazardous substances (e.g. protective gloves, safety goggles).
- Remove from the measuring chamber all inserted and screwed-in parts except the fastening pins.
- Pour the cleaning liquid into the cleaning bowl.
- Wet the measuring chamber with the cleaning liquid and allow it to act for some time depending on the degree of contamination (up to 24 hours for heavy contamination).
- Remove stubborn incrustations with the enclosed brushes.
- Rinse the measuring chamber with water.
- The cleaning liquid can be used several times. Pour the cleaning liquid back into the storage bottle using the funnel.
- Rinse the cleaning tray and brushes with water.
Instructions for maintenance of the measuring chamber of the Sycon analysers:
- Clean the sensor (Sycon 3000) and actuator with a soft cloth moistened with FIT 3000.
- Replace any wear parts when reassembling the measuring chamber.
- Grease the O-rings of the measuring chamber incl. the fixing pins.